UV-VIS Gel Imaging & Documentation System
Origin:China
Equipment introduction
FluorShot PRO II SC850 Gel Documentation System uses 5 megapixels professional high definition black/white CCD. Effective resolution is 2592*1944, SNR is 70db, pixel density is 12 bit, CCD acquisition system with two USB ports,
sensitivity of gel documentation in low-light level.
The system has UV automatic protection function. UV lamp automatically switches off when you are handling samples or it can enter to the glue cutting state and work
with special protecting glue cutting device for worker`s safety.
The top of the system can be easily open for changing filters and different fluorescent stains.
For standard 590nm filter you can use EB and GelRed stains, if you want to use other, please, contact us for changing hardware.
Main features
More models are available.
CE certification, more reliable.
Delicate computer, easy-moving, system noise is low.
Darkroom with mold production, enclosed camera, built-in acquisition system, dustproof.
Detection sensitivity: 20pg of double-stranded DNA.
Large area of UV transmission station, the biggest sample can be 21x26cm
Wavelength of UV Trans and Reflect: standard 312nm,254/365nm are available.
The most safety UV glue cutting device.
Specifications
Model: | FluorShot PRO II SC850 |
Acquisition system: | Professional high definition black/white CCD |
Effective resolution (pixels m) : | 2,592x 1,944 |
Pixel Size (um V*H): | 5.2*5.2 |
SNR: | 70db |
Pixel Density: | 12 bit |
Port: | USB 2.0 |
Lens: | Imported 6x motorized zoom F1.0, 2/3 inch high caliber penetrating lens |
Darkroom: | Mold production, enclosed camera, built-in acquisition system, dustproof and shockproof, UV |
Filter: | Standard 590 screw-type filter with coating (contact for special requirement) |
Operation interface: | With a panel or computer you can make full automatic control of zoom, focus, diaphragm, UV Trans and Reflect lamps. |
Display: | Computer |
UV Trans: | Standard 21*26cm with 312nm (254/365nm) |
UV Reflect: | Optional UV Reflect light source; used for paper chromatography and other Electrophoresis carrier which |
White Reflect: | Standard LED white light |
Fixed time functions: | |
Software: | Genescope fully control-analysis software with Chinese interface, Win2000/XP, image capture, edit, analysis and |
Power source: | 110/220V |
Dimension(mm): | 430*410*560 |
Weight: | 22kg |
Certification: | CE |